OCS1 Oxygen, Flow Rate Sensor Oxygen (±1.5%) 100ms response Flow Rate ±0-10 L/min (±0.1 L/min) UART, LCD & Analog outputs Operating temperature range –45 to +70ºC
|
|
Specification:
|
|
OCS1 |
|
1. Oxygen Concentrator sourced O2 mixture. (J1.4 open) 2. Medical/Hospital Ventilator Oxygen/Air mixer sourced O2 mixture. (J1.4 Set to GND) In circuit configurable to measure the Oxygen content in one of these gasses.
Powered by 8V to 35V / 7mA DC Power Supply, OCS1 provides UART & 0-5V O2 concentration & Flow Rate outputs. Thermal characteristics allow any Oxygen concentration measurement. No Flow Alarm output can serve to detect condition of No Flow.
|
|
Dimensions |
4.6” x 0.8” x 0.5” / 25g |
|
O2 source select |
J1.4 selects between O2 concentrator or 100% O2/Air mixer |
|
Outputs |
UART, LCD, Analog |
|
Analog O2 output |
0 to 5V DC Linearly proportional to 0—100% O2 |
|
Analog Flow Rate output |
0 to 5V DC Linearly proportional to 0—10 L/min |
|
Power Supply |
Unregulated 8V to 32V DC / 7 mA |
|
Field Calibration |
Oxygen & Flow rate DigiDISP controlled |
|
O2 Accuracy |
±1.5% @ –5 to +45°C; ±2% @ -45°C to –5°C, 45°C to +70°C |
|
O2 Range |
20.8% - 95.7%, 20.8% - 100% |
|
Flow Rate Accuracy |
±0.1 L/min |
|
Flow Rate Range |
±10 L/min |
|
Max pressure |
30 psig |
|
Thermal change response |
1 oC/min |
|
Thermal Gradientmax |
50 K/m |
|
Operating Temperature |
-45 to 70°C |
|
Survival Temperature |
-85 to +85°C |
|
RS232 transmit Rate |
200 bytes/sec @ 19200 bits/sec |
|
Calibration Retention |
More than 10 years |
|
Flow I/O |
Bi-directional via 1/8” male barbs |
|
O2 Response Time |
100 ms |
|
Flow rate Response Time |
10 ms |
|
USART Frame content |
O2 [%], Flow Rate [L/min] |
|
Flow I/O |
Bi-directional via 1/8” male barbs |
|
No Flow Alarm |
J1.3—Open Drain |
|
Long term display & save |
Via DigiTEST |
|
Converting to other units of Flow |
How to convert Volumetric Flow to other units of Flow Rate, such as gal/min, SLPM |